• Medientyp: E-Artikel
  • Titel: Impact of Different Metal Artifact Reduction Techniques on Attenuation Correction of Normal Organs in 18F-FDG-PET/CT
  • Beteiligte: Morawitz, Janna; Martin, Ole; Boos, Johannes; Sawicki, Lino M.; Wingendorf, Katrin; Sedlmair, Martin; Mamlins, Eduards; Antke, Christina; Antoch, Gerald; Schaarschmidt, Benedikt M.
  • Erschienen: MDPI AG, 2022
  • Erschienen in: Diagnostics
  • Sprache: Englisch
  • DOI: 10.3390/diagnostics12020375
  • ISSN: 2075-4418
  • Schlagwörter: Clinical Biochemistry
  • Entstehung:
  • Anmerkungen:
  • Beschreibung: <jats:p>Purpose: To evaluate the impact of different metal artifact reduction algorithms on Hounsfield units (HU) and the standardized uptake value (SUV) in normal organs in patients with different metal implants. Methods: This study prospectively included 66 patients (mean age of 66.02 ± 13.1 years) with 87 different metal implants. CT image reconstructions were performed using weighted filtered back projection (WFBP) as the standard method, metal artifact reduction in image space (MARIS), and an iterative metal artifacts reduction (iMAR) algorithm for large implants. These datasets were used for PET attenuation correction. HU and SUV measurements were performed in nine predefined anatomical locations: liver, lower lung lobes, descending aorta, thoracic vertebral body, autochthonous back muscles, pectoral muscles, and internal jugular vein. Differences between HU and SUV measurements were compared using paired t-tests. The significance level was determined as p = 0.017 using Bonferroni correction. Results: No significant differences were observed between reconstructed images using iMAR and WFBP concerning HU and SUV measurements in liver (HU: p = 0.055; SUVmax: p = 0.586), lung (HU: p = 0.276; SUVmax: p = 1.0 for the right side and HU: p = 0.630; SUVmax: p = 0.109 for the left side), descending aorta (HU: p = 0.333; SUVmax: p = 0.083), thoracic vertebral body (HU: p = 0.725; SUVmax: p = 0.392), autochthonous back muscles (HU: p = 0.281; SUVmax: p = 0.839), pectoral muscles (HU: p = 0.481; SUVmax: p = 0.277 for the right side and HU: p = 0.313; SUVmax: p = 0.859 for the left side), or the internal jugular vein (HU: p = 0.343; SUVmax: p = 0.194). Conclusion: Metal artifact reduction algorithms such as iMAR do not alter the data information of normal organs not affected by artifacts.</jats:p>
  • Zugangsstatus: Freier Zugang